Cooling scheme | Water-cooling type |
---|---|
Cooling capacity (W) | 5900 / 6400 |
Heating capacity / rapid heating(W) | 0.0 |
Heating capacity / controlled heating(W) | 6000 |
Circulated water | Pure water or distilled water (1μs/cm or more), ethylene glycol solution (concentration: 30% max.) |
Temperature stability (℃) | ±0.10 (High precision mode) |
Use Area Temperature (℃) | 5~40 |
Water setting temperature range (℃) | -10 to 80 |
Circulating pump/Rated output (W) | 750 |
Circulating pump/Flow rate (ℓ/min) | 35/40 |
Circulating pump/Head (m) | 24/35 |
Maximum tank capacity (ℓ) | 38 |
Cooling water / Temperature range (℃) | 5~40 |
Cooling water / Pressure (Mpa) | 0.1~0.5 |
Cooling water / Flow rate (ℓ/min) | 25 |
Dimensions (mm) | H1250×W440×D800 |
Weight (kg) | 137 |
Rated Voltage (V) | 3-phase AC200 |
Power Consumption/MIN (A) | 2900/3350 |
Power Consumption/MAX (W) | 8900/9350 |
Current Consumption/MIN (A) | 10.5/11.2 |
Current Consumption /MAX (A) | 30.5/31.2 |
Voltage source capacity (kVA) | 15 |
Refrigerator | R407C 1100g |
Noise (dB) | 55/55 |
Safety circuit / Compressor | Temperature protector (135℃ OFF), overcurrent relay |
Safety circuit / Refrigeration cycle | Pressure switch |
Safety circuit / Circulating water circuit | Water pressure sensor, float switches for water shortage/water supply/water full detection |
Safety circuit/Pump | Overcurrent relay |
Safety circuit / Fan motor | Overcurrent relay |
Safety circuit / Heater | Overheat prevention thermostat (90℃ OFF), electric current fuse Alarm output |
Input/Output / Alarm output | a-contact, no voltage (250 VAC/2 A/resistive load) |
Input/Output / Water temperature upper/lower limit output | a-contact, no voltage (250 VAC/2 A/resistive load) |
Input/Output / Water supply valve/operation output | a-contact, no voltage (250 VAC/2 A/resistive load) |
Input/Output / External control input | No-voltage input |
Input/Output / Emergency stop input | No-voltage input |
Input/Output / RS-232C | a-contact, no voltage (250 VAC/2 A/resistive load) |
Input/Output / External temperature sensor input | Pt100 |
Safety measures | Compressor conditioning timer, closing assist pump, fixing adjuster, double safety pan structure |
Functions | Circulating water pressure display, heater power display, pump operating time display, freeze-proof operation, digital filter, password setting, automatic tuning, ON/OFF timer, key-lock, error history memory |